
Inkqubo yokusebenza ekrelekrele; Ukusebenza ngokulula ngeealam ezibonakalayo kunye nezandi; Ujongano lwenkonzo/ugcino lweenjongo ezininzi; Iprofayili: i-sodium concentration kunye ne-UF curve.
I-W-T6008S iqinisekisa ukhuseleko kunye nokusebenza kakuhle ngexesha le-dialysis, inikezela ngonyango lwe-dialysis olukhululekile, olungasebenzisa: I-HDF ye-intanethi, i-HD kunye ne-On-line HF.
kwi-intanethi HDF
Igumbi lokulinganisa elivaliweyo elivaliweyo, ulawulo oluchanekileyo lwe-ultrafiltration yokuphelelwa ngamanzi emzimbeni;Isitshixo esinye sesantya esisezantsi se-ultrafiltration: inokuseta isantya esiphantsi se-UF, isantya esiphantsi se-UF ixesha lokusebenza, ukubuyela kwisantya esiqhelekileyo se-UF ngokuzenzekelayo emva kokubulawa; Inkxaso ye-UF eyodwa, inokuguqula ixesha elisetyenzisiweyo kunye nevolumu ye-UF ngokusekwe kwimfuno ngelixa i-UF yodwa.
Isitshixo esinye se-dialyzer priming+ umsebenzi
Inokuseta ixesha lokuqala, i-priming dehydration volume ekusetyenzisweni okusebenzayo kokusasazwa kunye nendlela yokuhambisa ukuphucula isiphumo sokuqala se-bloodlines kunye ne-dialyzer kunye nokuphucula ukwanela kwe-dialysis.
Ubukrelekrele obuzenzekelayo bokubulala iintsholongwane kunye nenkqubo yokucoca
Inokuthintela ngokufanelekileyo ukubekwa kwekhalsiyam kunye neprotheyini kumbhobho womatshini, ngokungeyomfuneko ukusebenzisa i-sodium hypochlorite ukususa iprotein ethintela ukwenzakala kwabasebenzi bezonyango ngexesha lokusetyenziswa kwe-sodium hypochlorite.
Umsebenzi ongundoqo wokukhupha amanzi
Isixhobo esiluncedo nesisebenzayo sokukhupha amanzi, susa ngokuzenzekelayo ulwelo lwenkunkuma kwi-bloodline kunye ne-dialyzer emva konyango lwe-dialysis, oluthintela ulwelo lwenkunkuma ukuba lungachitheki emhlabeni xa kuchithwa umbhobho, ngokufanelekileyo gcina indawo yonyango icocekile kwaye inciphise ulawulo kunye neendleko zothutho lwenkunkuma yonyango.
Inkqubo ye-alamu yesixhobo esikrelekrele seHemodialysis
Ingxelo yembali ye-alam kunye nokubulala iintsholongwane
15 intshi LCD touch screen
Kt/V uvandlakanyo
Ukusetwa kweparamitha yesodium kunye ne-UF ngokwezifiso ngokusekwe kwimeko yokwenyani yonyango lwezigulane, olulungele unyango olulolwakho lwezonyango, izigulana ziya kuziva zikhululekile ngakumbi ngexesha lokucofa igazi kwaye zinciphise izehlo zokusabela okungathandekiyo okuqhelekileyo.
| Ubungakanani &Ubunzima | |
| Ubungakanani | 380mmx400x1380mm (L*W*H) |
| Net Weight malunga. | 88KG |
| Ubunzima obuGqibeleleyo malunga. | malunga ne-100KG |
| Ubungakanani bepakethi malunga. | 650×690×1581mm (L x W x H) |
| Ukunikezwa Amandla | |
| AC220V, 50Hz/60Hz, 10A | |
| Igalelo lamandla | 1500W |
| Back-up ibhetri | Imizuzu engama-30 |
| Imeko yokusebenza | |
| Uxinzelelo lwegalelo lamanzi | 0.1Mpa ~ 0.6Mpa, 15P.SI~60P.SI |
| Ubushushu begalelo lamanzi | 5℃~30℃ |
| Ubushushu bendawo yokusebenza | 10℃~30℃ kumanzi anxulumeneyo ≦70% |
| Ixabiso le-UF | |
| Uluhlu lokuhamba | 0ml/h~4000ml/h |
| Umlinganiselo wesigqibo | 1ml |
| Ukuchaneka | ±30ml/h |
| Impompo yegazi kunye nempompo yokutshintsha | |
| Uluhlu lokuhamba kwempompo yegazi | 10ml/min~600ml/min (ububanzi: 8mm okanye 6mm) |
| Uluhlu lokuhamba kwempompo yokutshintsha | 10ml/min~300ml/min (ububanzi 8mm okanye 6mm) |
| Umlinganiselo wesigqibo | 0.1ml |
| Ukuchaneka | ±10ml okanye i-10% yokufunda |
| Impompo yeHeparin | |
| Ubungakanani besirinji | 20, 30, 50ml |
| Uluhlu lokuhamba | 0ml/h~10ml/h |
| Umlinganiselo wesigqibo | 0.1ml |
| Ukuchaneka | ±5% |
| Inkqubo yokubek'esweni kunye nokuseta iAlarm | |
| Uxinzelelo lwe-venous | -180mmHg ~ +600mmHg, ± 10mmHg |
| Uxinzelelo lwe-arterial | -380mmHg ~ +400mmHg, ±10mmHg |
| TMP | -180mmHg ~ +600mmHg, ±20mmHg |
| Ubushushu be-Dialysate | uluhlu olusetwe kwangaphambili 34.0℃~39.0℃ |
| Ukuhamba kwe-dialysate | Ngaphantsi kwe-800 ml/min (Inokulungiswa) |
| Uluhlu lokuhamba endaweni | 0-28 L/H (kwi-intanethi HDF) |
| Ukubona ukuvuza kwegazi | Photo chromic alarm xa erythrocyte umthamo ethile 0.32±0.02 okanye ukuvuza umthamo igazi ilingana okanye ngaphezu 1ml ilitha nganye dialysate. |
| Ukufunyanwa kwebhamuza | I-Ultrasonic, i-Alarm xa umthamo weqamza lomoya omnye ungaphezulu kwe-200μl kwi-200ml/min ukuhamba kwegazi |
| Ukuqhuba | I-acoustic-optic |
| Ukubulala iintsholongwane/ukucoca | |
| 1. Ukubulala iintsholongwane ezishushu | |
| Ixesha: imizuzu engama-30; Ubushushu: malunga ne-80 ℃, kwinqanaba lokuhamba kwe-500ml / min; | |
| 2. Ukubulala iintsholongwane ngamachiza | |
| Ixesha: 30minutes, Ubushushu: malunga ne-36 ℃ ~ 50 ℃, kwinqanaba lokuhamba 500ml / min; | |
| 3. I-Chemical disinfection ngobushushu | |
| Ixesha: 45minutes, Ubushushu: malunga ne-36 ℃ ~ 80 ℃, kwinqanaba lokuhamba kwe-50ml / min; | |
| 4. Hlanza | |
| Ixesha: 10minutes, Ubushushu: malunga ne-37 ℃, kwinqanaba lokuhamba kwe-800ml / min; | |
| Indawo yogcino | |
| Ubushushu bokugcina kufuneka bube phakathi kwe-5℃~40℃, kwindawo yokufuma ≦80% | |
| Umsebenzi | |
| I-HDF, i-BPM ye-intanethi, i-Bi-cart kunye ne-2 pcs endotoxin filters | |
